MAUDE Adverse Event Report: WRIGHTS LANE : SYNTHES USA PRODUCTS LLC TI MATRIXRIB I-PLATE WIDE 2.8MM THICK/11 HOLES - STERILE; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE : SYNTHES USA PRODUCTS LLC TI MATRIXRIB I-PLATE WIDE 2.8MM THICK/11 HOLES - STERILE; PLATE, FIXATION, BONE Back to Search Results
Model Number 04.501.104S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/30/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient weight not available for reporting.Complainant part is not expected to be returned for manufacturer review/investigation.Device malfunctioned intra-operatively and was not implanted / explanted.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent a surgical procedure on (b)(6) 2017 as treatment for a fractured rib secondary to being involved in a motor vehicle accident.During the procedure, the surgeon was bending an 11-hole matrixrib plate at the back table and it broke in half.The plate fragments were retrieved.There was no harm reported to the patient.Another plate was bent without incident and implanted in the patient.These events resulted in a 2-minute surgical delay.No harm was reported to the patient.The procedure was successfully completed.The patient outcome was reported stable.Concomitant devices reported: bending pliers (part number unknown, lot number unknown, quantity 1).This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Year of birth is not (b)(6).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Patient date of birth is not known.Dhr review was completed.Part no.: 04.501.104s, lot no.: l193252, manufacturing location: (b)(4), release to warehouse date: 09.Jan.2017, expiry date: 30.Nov.2026.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Corrected data: date of event, concomitant device therapy date.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI MATRIXRIB I-PLATE WIDE 2.8MM THICK/11 HOLES - STERILE
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE : SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7156983
Report Number2939274-2018-50005
Device Sequence Number18867196
Product Code HRS
UDI-Device Identifier10887587073330
UDI-Public(01)10887587073330(17)261130(10)L193252
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K161590
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.501.104S
Device Catalogue Number04.501.104S
Device Lot NumberL193252
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/04/2017
Supplement Date Received by Manufacturer01/12/2018
01/12/2018
Initial Report FDA Received Date01/02/2018
Supplement Report FDA Received Date01/12/2018
01/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
-
-