MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.4MM LCP® STRAIGHT WRIST PLATE 170MM; PLATE,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.4MM LCP® STRAIGHT WRIST PLATE 170MM; PLATE,FIXATION,BONE Back to Search Results
Model Number SD242.003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional narrative: (b)(4).Date of postoperative plate breakage is unknown.Complainant device is not expected to be returned for manufacturer review/investigation.(b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that initially a patient underwent for an open reduction internal fixation (orif) of distal radius on an unknown date and was implanted with wrist plate.The plate broke postoperative due to non-compliance of patient.The patient received extensor tendon ruptures due to hardware failure.Hardware removal of broken plate was performed on an unknown date.Surgery was completed successfully with no surgical delay reported.Patient outcome reported as stable following surgery.This report is for one (1) 2.4mm lcp® straight wrist plate 170mm.(b)(4).
 
Additional Manufacturer Narrative
Device history record review.Part number: sd242.003 lot number: h271201.Part manufacture date: 27-jan-2017 manufacturing location: (b)(4) part expiration date: n/a nonconformance noted: n/a dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 2.4mm lcp straight wrist plate 170mm product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no nonconformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.4MM LCP® STRAIGHT WRIST PLATE 170MM
Common Device Name
PLATE,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7157337
Report Number2939274-2018-50011
Device Sequence Number111808
Product Code HRS
UDI-Device Identifier10886982003638
UDI-Public(01)10886982003638(10)H271201
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K040588
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSD242.003
Device Catalogue NumberSD242.003
Device Lot NumberH271201
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/04/2017
Supplement Date Received by Manufacturer01/31/2018
Initial Report FDA Received Date01/02/2018
Supplement Report FDA Received Date02/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight72
-
-