ALLERGAN (COSTA RICA) STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133MV-13-T
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/04/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: avoiding damage during surgery.Extreme care should be taken to avoid damage to the tissue expander during surgery.Possible sources of damage include sharp surgical instruments such as scalpels and needles used during the initial surgery, subsequent filling, or hematoma/fluid evacuation.A sterile back-up tissue expander must be readily available at the time of surgery in the event that damage occurs.Products must be carefully inspected for leaks or nicks prior to use.Do not attempt to repair damaged products.
 
Event or Problem Description
Healthcare professional reported a "leak" with a tissue expander.Device was not implanted.Surgery was completed with a backup expander.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device not identified observations.A leak test was performed which did not identify openings on the device.Microscopic analysis was performed which did not identify any additional observations.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.No openings on the device and no damage on valve; broken device was not confirmed.
 
Event or Problem Description
Healthcare professional reported a "leak" with a tissue expander.Device was not implanted.Surgery was completed with a backup expander.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key7157598
Report Number9617229-2017-05671
Device Sequence Number13653468
Product Code LCJ
UDI-Device Identifier10888628004429
UDI-Public10888628004429
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/25/2021
Device Catalogue Number133MV-13-T
Device Lot Number3118794
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2018
Initial Date Received by Manufacturer 12/12/2017
Supplement Date Received by Manufacturer01/31/2018
Initial Report FDA Received Date01/02/2018
Supplement Report FDA Received Date02/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-