| Model Number |
6996SQ41 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Over-Sensing (1438); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/03/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported there was a lead warning, the lead displayed variability in defibrillation impedance, there was an increase in the right ventricular lead impedance and it was high, there was a high rate of non-sustained episodes, and the sensing integrity count was noted.Intervention is planned; the lead will be capped and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was later reported the lead displayed t wave over-sensing and fracture.
|
| |
|