| Model Number |
FPRPR3560 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the testing of 2 icesphere needles, one needle shaft started to collect frost immediately.Needle was replaced to continue with testing.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The needle was returned for investigation.The manufacturing records were reviewed and no related deviation or concerns were found.The reported frost on the needle shaft was confirmed as the needle failed investigative testing.The needle was disassembled and inspected.No visible soldering gaps or voids were found.Microscopic inspection of the vacuum sleeve was conducted and corrosive pits or soldering flux residuals were identified on the external surface of the vacuum sleeve.A bubble test of the vacuum sleeve was conducted and a leak was identified at the top end of the sleeve.The leaked zone was further investigated under microscope and a hole on the external tube crimping area was observed.The root cause of the hole development is not known.Based on the investigation results, the customer complaint was verified and the root cause was determined to be a component failure.No relationship was identified between the needle assembly processes and the vacuum loss phenomena.
|
| |
|