Model Number FPRPR3560
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/01/2017
Type of Reportable Event Malfunction
Event or Problem Description
During the testing of 2 icesphere needles, one needle shaft started to collect frost immediately.Needle was replaced to continue with testing.There was no patient involvement reported.
 
Additional Manufacturer Narrative
The needle was returned for investigation.The manufacturing records were reviewed and no related deviation or concerns were found.The reported frost on the needle shaft was confirmed as the needle failed investigative testing.The needle was disassembled and inspected.No visible soldering gaps or voids were found.Microscopic inspection of the vacuum sleeve was conducted and corrosive pits or soldering flux residuals were identified on the external surface of the vacuum sleeve.A bubble test of the vacuum sleeve was conducted and a leak was identified at the top end of the sleeve.The leaked zone was further investigated under microscope and a hole on the external tube crimping area was observed.The root cause of the hole development is not known.Based on the investigation results, the customer complaint was verified and the root cause was determined to be a component failure.No relationship was identified between the needle assembly processes and the vacuum loss phenomena.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICESPHERE 1.5 90 CRYOABLATION NEEDLE
Common Device Name
CRYOABLATION NEEDLE
Manufacturer (Section D)
GALIL MEDICAL LTD
1 tavor building, p.o. box 224
industrial park
yokneam,, israel 20692
IS  20692
MDR Report Key7158230
Report Number9616793-2018-00001
Device Sequence Number949052
Product Code GEH
UDI-Device Identifier07290104830945
UDI-Public07290104830945
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K141485
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/30/2018
Device Model NumberFPRPR3560
Device Lot NumberB8749
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/04/2017
Supplement Date Received by Manufacturer01/15/2018
Initial Report FDA Received Date01/02/2018
Supplement Report FDA Received Date01/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-