| Model Number |
VLOCM0014 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/13/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, when both of the sutures were taken out of the package, the strand was loose from the needle and the thread broke.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned noted one suture on a retainer.The needle was observed to be detached from the suture; the needle wasn't returned.The loop is intact.The retainer was inspected with no abnormalities.The retainer was inspected with no abnormalities.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Unfortunately, because no sealed samples were received, a tensile test and/or needle attachment test could not be performed.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|