• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522LNAH
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 12/21/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer called and reported that they got hospitalized due to low blood glucose on (b)(6) 2017 with blood glucose of 20 mg/dl at the time of the incident.The customer was at 37 mg/dl at the time of admission, and 176 mg/dl at the time of the call.The customer was given glucagon shot to treat.The customer experienced symptoms such as a seizure and loss of consciousness.The customer was wearing the insulin pump during the incident.Troubleshooting was not completed as the customer declined.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device uploaded properly using carelink.Device had minor scratched display window and cracked reservoir tube lip.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7159079
Report Number3004209178-2018-40226
Device Sequence Number19184255
Product Code OYC
UDI-Device Identifier00613994597212
UDI-Public(01)00613994597212
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 10/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-522LNAH
Device Catalogue NumberMMT-522LNAH
Device Lot NumberA2522LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Initial Date Received by Manufacturer 12/24/2017
Supplement Date Received by Manufacturer10/25/2019
Initial Report FDA Received Date01/02/2018
Supplement Report FDA Received Date10/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age28 YR
Patient Weight106
-
-