• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL, ASD, INC. PORTEX® BIVONA® TTS¿ TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL, ASD, INC. PORTEX® BIVONA® TTS¿ TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Catalog Number FT16LN35NGF124N
Medical Device Problem Codes Air Leak (1008); Calcified (1077); Leak/Splash (1354); Loose or Intermittent Connection (1371); Component Missing (2306); Failure to Align (2522); Naturally Worn (2988)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/21/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a portex® bivona® tts¿ custom tracheostomy tube was leaking from the cuff.There was no patient injury.
 
Additional Manufacturer Narrative
One pneupac® parapac® ventilator was received with a battery, oxygen input hose, air input hose, air gas supply indicator shroud ring missing, and eyeball window not correctly in position.Upon visual inspection, damage was noted to the casing and exhalation valve.There was also noted alignment of the peep/ cpap control dial and the detent calibrated mark.Two screws were observed to be loose inside the unit after the unit was opened for further inspection.The screws are reported to hold the gas supply indicator in position.Functional testing was performed and revealed that the relief valve was below specification for different positions (40, 50, 60, & 70).However, noted to not have contributed to the reported fault.Non-return valve assemblies were removed from oscillator block allowing for further investigation finding "swarf" on the non-return valve.The reported swarf was noted to be preventing the valve disc from sealing causing a leak within the unit.Although the reported fault could not be duplicated testing, the observations noted on non-return valve could have initiated a leak causing the increase in inspiratory time, leading to oscillator stall; which could cause the unit to fail.Reason of the swarf is unknown, oscillator block inspected and no further evidence of swarf was noted.Other faults and damages identified would not have contributed towards the reported fault.No adverse trends have been identified relating to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® BIVONA® TTS¿ TRACHEOSTOMY TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL, ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7160365
Report Number3012307300-2018-02431
Device Sequence Number19381661
Product Code JOH
UDI-Device Identifier15019517090161
UDI-Public15019517090161
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/10/2023
Device Catalogue NumberFT16LN35NGF124N
Device Lot NumberDS016156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2018
Supplement Date Received by Manufacturer05/25/2018
Initial Report FDA Received Date01/03/2018
Supplement Report FDA Received Date01/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-