| Model Number |
486619100 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; height adjuster tip broke while adjusting screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, product history review, complaint history review, risk assessment; result: the reported event was confirmed fractured via visual inspection of the returned product.All fragments were removed and there was no reported delay.Manufacturing files were reviewed and no issues were found.The stryker rep confirmed that the instrument was being used correctly during surgery.The likely root causes of the deformation was instrument age, normal wear and tear.
|
| |
|
Event or Problem Description
|
|
It was reported that; height adjuster tip broke while adjusting screw.There are no reports of adverse consequences to the patient.
|
| |
|