| Model Number |
6996SQ41 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/28/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported an alert triggered due to a sudden increase in the superior vena cava (svc) coil, which occurred following a lead fracture.A radiograph and imaging under fluoroscopy showed a coil fracture.It was also reported the svc impedance was high and the patient described a "poking" feeling around the pocket.A partial explant was performed; the portion of the lead connected to the device was removed and the distal portion capped.The lead was replaced, and no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|