Model Number MT22949
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had intermittent vibration.No additional event or patient information is available.No product was provided for evaluation.The complaint confirmation of the receiver having intermittent vibration could not be determined.A root cause could not be determined.
 
Additional Manufacturer Narrative
Sr (b)(4).
 
Event or Problem Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and the and there was no failure detected.A manual test was performed and there was vibration omitting from the device.A review of the downloaded data log did not find any errors related to the customer complaint.The reported event of an intermittent vibration was not confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key7161188
Report Number3004753838-2018-00907
Device Sequence Number13827218
Product Code PQF
UDI-Device Identifier30386270000324
UDI-Public30386270000324
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22949
Device Catalogue NumberSTR-SD-001
Device Lot Number5231267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2017
Supplement Date Received by Manufacturer02/06/2018
Initial Report FDA Received Date01/03/2018
Supplement Report FDA Received Date03/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-