• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE40

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE40 Back to Search Results
Model Number 400-01-221
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/12/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having dislocated.The surgeon "switched" from bipolar to total hip.He used an orthority liner.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a dislocation.The previous surgery and the revision detailed in this investigation occurred 9.5 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a dislocation.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a dislocation.The previous surgery and the revision detailed in this investigation occurred 9.5 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.The root cause of this complaint was a revision surgery due to a dislocation.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR HIP
Common Device Name
SHELL, BIPOLAR, MODULAR, ENCORE, SIZE40
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7161639
Report Number1644408-2018-00018
Device Sequence Number1671213
Product Code KWY
UDI-Device Identifier00888912075626
UDI-Public(01)00888912075626
Combination Product (Y/N)N
Initial Reporter StatePR
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/25/2022
Device Model Number400-01-221
Device Catalogue Number412-02-040
Device Lot Number982B1020
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/12/2017
Supplement Date Received by Manufacturer02/13/2018
02/21/2018
Initial Report FDA Received Date01/03/2018
Supplement Report FDA Received Date02/15/2018
02/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
-
-