| Model Number |
42632-05 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned to the manufacturer for investigation.It has not yet been received.
|
| |
|
Event or Problem Description
|
|
The event involved a customer report of a transpac iv monitoring kit that disconnected at the connection of the device.It was reported that there was suspicion of missing solvent to glue it properly.No adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One (1) used 'partial set', transpac iv mgt.Kit was received for evaluation with no tubing separation.The kit was primed and pressure tested.During priming, a leak was observed between the male luer and the 60" administration tubing.The administration tubing was found to separate during handling of the product, confirming a separation between the administration tubing and the male luer.This was caused by insufficient solvent around the tubing.
|
| |
|