The recipient reportedly experienced loss of sound.External equipment was exchanged and programming adjustments were made, however, issue was not resolved.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
The external visual inspection revealed the electrode was cut prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed cut electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.The internal visual inspection noted silver migration shorting across electrical components.This device had moisture that exceeded the residual gas analysis test limit.Based on an assessment of the residual gas analysis and dye penetrant data, it is determined that this device was non-hermetic, and that the root cause of the excessive moisture was a leak through the feed thru seals.A corrective action was implemented.This is the final report.
|