Model Number 3650
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 12/15/2017
Event Type  malfunction  
Event Description
It was reported that during a device preparation, the programmer was noted to shock the users.Whenever the programmer was used, it would shock the user.There was no patient involved.
 
Manufacturer Narrative
The field complaint of static electricity shock was verified.During analysis, paper was loaded into the printer and test pages were printed.The paper generated static electricity once it passed through the roller of the printer.Investigation revealed that the moisture content of the paper being too low was the cause of the complaint.It was most likely stored outside of the plastic packaging in a dry environment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7162515
MDR Text Key96354488
Report Number2938836-2018-00012
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Device Catalogue Number3650
Other Device ID Number05414734500661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/15/2017
Initial Date FDA Received01/03/2018
Supplement Dates Manufacturer Received02/22/2018
02/21/2019
Supplement Dates FDA Received03/12/2018
02/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-