| Model Number |
A28-28/C95V |
| Medical Device Problem Codes |
Failure To Adhere Or Bond (1031); Leak/Splash (1354); Migration or Expulsion of Device (1395); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Failure of Implant (1924); Pain (1994); Rupture (2208)
|
| Date of Event |
12/05/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in the event will not be returned for evaluation, they remain implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with a bifurcated stent and two infrarenal aortic extensions.The patient came in emergently on (b)(6) 2017 with severe abdominal pain.The physician identified a migration of the proximal extension and a type 1a endoleak.The physician elected to implant two infrarenal aortic extensions and two non-endologix renal stents to complete a chimney procedure.The patient is reported to be doing well post procedure.There have been no additional adverse events reported for this patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information received at the completion of the clinical evaluation, there were substantial reported evidence that supported the following case event.Clinical evaluation was able to confirmed that patient had a type 1a endoleak and stent migration.In addition, patient also had a sac growth, and a rupture.A secondary procedure was performed on 12/05/2017 to resolve issue.Cumulative knowledge informed by past assessments of similar complaints was applied to a review of the available medical information.The most likely cause of the proximal loss of seal was the placement of the cuff in an off-label neck anatomy (intentional user error) which resulted in stent migration and a contained rupture.Procedure related harms included the type ii endoleak.The final patient disposition could not be ascertained due to lack of medical information surrounding the secondary endovascular procedure.Reportedly, the patient was in stable condition post the secondary procedure.To date there has been no reports of further negative patient sequelae.In addition, the review of manufacturing lot confirmed all devices met specifications prior to release.The devices remain implanted in the patient and were not returned and no evaluation was completed.These types of events will be monitored and trended as part of the quality system.
|
| |
|