• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACCLARENT, INC. RELIEVA SCOUT® MULTI-SINUS DILATION SYSTEM 6X24MM (5 PK); INSTRUMENT, ENT MANUAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACCLARENT, INC. RELIEVA SCOUT® MULTI-SINUS DILATION SYSTEM 6X24MM (5 PK); INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Catalog Number SCT0624MFS
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cerebrospinal Fluid Leakage (1772)
Date of Event 12/13/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The patient identifier is not known.The patient¿s age / date of birth is not known.The patient's weight is not known.A review of the lot history record the subject device (lot# 160809a-pc) did not present any issues during the manufacturing or inspection process that can be related to the reported complaint.The acclarent products used during the procedure were said to have functioned as expected.No malfunctions were reported with any acclarent devices used during this procedure.No acclarent devices are available to be returned for evaluation as they were discarded by the user facility.Additional information will be submitted within 30 days of receipt.(b)(4).
 
Event or Problem Description
Acclarent was informed on (b)(6) 2017, of an event that occurred on (b)(6) 2017, involving a relieva scout® multi-sinus dilation system 6 x 24 mm (lot#: 160809a-pc).During a primary functional endoscopic sinus surgery (fess) procedure, the patient developed a cerebrospinal fluid (csf) leak after the first inflation at 12 atm in the right frontal sinus.The csf leak was patched with a mucosal graft and duraseal and the fess procedure was completed as planned.The patient stayed overnight in the hospital after the procedure and was reported to be doing well after being discharged from the hospital the next day.It was reported that the physician felt that the csf leak may have been caused by his probing with a seeker more than the balloon inflation itself.However, the cause was not determined after the fact.It was also reported that this was the first time the physician had used the rigid non-acclarent device.The relieva scout® multi-sinus was reported to have functioned as expected.There were no reported malfunctions with any acclarent devices used during this procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIEVA SCOUT® MULTI-SINUS DILATION SYSTEM 6X24MM (5 PK)
Common Device Name
INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
ACCLARENT, INC.
33 technology drive
irvine 92618
MDR Report Key7162769
Report Number3005172759-2017-00060
Device Sequence Number63667
Product Code LRC
UDI-Device Identifier10705031238039
UDI-Public(01)10705031238039(11)160809(17)170809(10)160809A-PC
Combination Product (Y/N)N
PMA/510(K) Number
K120280
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSCT0624MFS
Device Lot Number160809A-PC
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2017
Initial Report FDA Received Date01/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-