| Model Number |
72200873SR |
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by customer that device presented a short circuit with the smell of smoke, before the case and no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Product failed functional testing with a short circuit error.Complaint of "short circuit" error message and burnt smell was confirmed.Cause of errors is a shorted electronic component on the main digital pcb.Burnt smell was caused by burnt resistor on main pcb.Product passed functional testing with a known good main pcb installed.
|
| |
|