• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Blurred Vision (2137); Visual Impairment (2138)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown, information not provided.Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.If explanted, give date: not applicable, as the lens remains implanted.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zlb00 19.5 diopter intraocular lens (iol) was implanted in the patient's left eye (os) on (b)(6) 2017.The patient reported his vision was great following the implantation procedure even though he developed a minor infection which was treated by medication.However, the patient experienced visual difficulties after a yttrium-aluminum garnet (yag) was performed in (b)(6) 2017.The patient complained of not being able to read up close, distance vision was blurry, seeing halos at night, dizziness, and headaches.Reportedly, the doctor explained to the patient these complaints were due to dry eyes, which the patient had treated for with plugs prior to cataract surgery.The patient still has dry eyes.Also, when the patient asked why the yag procedure was performed, the doctor's reason was because it was routine for their office to perform this procedure at a patient's one month post-operative visit.No additional information was provided.The patient had bilateral lenses implanted.This report will capture the lens implanted in the patient's left eye.A separate report will be filed for the right eye.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the complaint device was not returned for evaluation.The reported complaint cannot be verified.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to the specifications.A search of complaints was performed on 15-dec-2017.The search revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7162997
Report Number9614546-2018-00018
Device Sequence Number1340377
Product Code MFK
UDI-Device Identifier05050474560666
UDI-Public(01)05050474560666(17)200321
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date03/21/2020
Device Model NumberZLB00
Device Catalogue NumberZLB00U0195
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/06/2017
Supplement Date Received by Manufacturer01/14/2018
Initial Report FDA Received Date01/03/2018
Supplement Report FDA Received Date02/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-