• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4002C0312
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pc400 introducer sheath was kinked approximately 4.0 cm from its distal end.Conclusions: evaluation of the returned pc400 revealed a kinked introducer sheath.This damage was likely due to improper handling during preparation of the pc400.If this introducer sheath becomes kinked it is likely resistance will be experienced while advancing the embolization coil through the kinked region.If the kink is severe enough the device may not be able to advance out of its sheath.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the inferior mesenteric artery (ima) using penumbra coil 400s (pc400s).During the procedure, the physician experienced resistance and was unable to advance a pc400 out of its introducer sheath and into the lantern delivery microcatheter (lantern), and therefore the pc400 was removed.The procedure was then completed using 6 additional pc400s and the same lantern.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7163086
Report Number3005168196-2018-00013
Device Sequence Number19279106
Product Code HCG
UDI-Device Identifier00814548013367
UDI-Public00814548013367
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/28/2024
Device Catalogue Number4002C0312
Device Lot NumberF71843
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2017
Initial Report FDA Received Date01/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
-
-