Exemption number e2016032.(b)(4).After investigation the event for this pr# is no longer reportable.(b)(4).Summary of investigational findings: the reported information states that the physician observed a hole/crack in the stopcock of the device and subsequently leakage when flusing it.The hole was glued and the stent graft was placed and deployed in the patient.Hereafter, the connecting tube began leaking again and the procedure was completed quickly.Blood loss was reported as a consequence of the damage of the connecting tube.No adverse effects were reported.The evaluation of the returned product found a 4 mm hole on the connecting tube.Even though the damage on the connecting tube was found prior to use, the physician chose to glue the hole and use the device.As per ifu, the product has to be inspected prior to use and if damage has occured, it is not recommended to use the product and instead, it has to be returned to cook.It is not possible to determine the actual root cause because of limited information, however, since the work order of the device showed no evidence to suggest a manufacturing out of specification, the most likely cause is that the damage occurred during shipping of the device.It is not possible to determine the acutal root cause because of limited information, however, since the work order of the device showed no evidence to suggest a manufacturing out of specification, the most likely cause is that the damage occured during shipping of the deivce.Cook medical will continue to monitor for similar events.
|