Catalog Number 4922045
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-op diagnosis: posterior scoliosis procedure: oblique lumbar interbody fusion(olif), lateral lumbar interbody fusion(xlif) levels: l2/3/4/5 it was reported that during surgery, the cage was allegedly reported broken.After the cage was inserted using an inserter that has sleeves, the cage was removed using a remover to adjust its position.When the cage was attached to the inserter to reinsert, the force grasping the cage was found to be weak.The cage was replaced.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: the threads on the backside of the implant have been stripped out.Witness marks on the backside of the implant indicate the implant was tightened fully.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLYDESDALE PTC SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf 94469
MDR Report Key7163222
Report Number1030489-2018-00009
Device Sequence Number13827273
Product Code MAX
UDI-Device Identifier00643169190801
UDI-Public00643169190801
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133205
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/23/2023
Device Catalogue Number4922045
Device Lot Number35AF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Initial Date Received by Manufacturer 12/05/2017
Supplement Date Received by Manufacturer04/02/2018
Initial Report FDA Received Date01/04/2018
Supplement Report FDA Received Date04/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
-
-