MDT KYPHON NEUCHATEL MFG KYPHON EXPRESS INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number KE152
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, intra-op, the bone tamp ruptured during inflation and would not fit back into the cannula when retrieval was attempted.Product came in contact with the patient.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluation: the balloon has been caught on the edge of the tube and torn.This damage was caused by the balloon being cut by sharp bone during inflation.Since the seal has been lost a vacuum cannot be created to fully deflate the balloon.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON EXPRESS INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel
CH 
MDR Report Key7163229
Report Number2953769-2018-00001
Device Sequence Number63877
Product Code HRX
UDI-Device Identifier00643169143777
UDI-Public00643169143777
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K123771
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberKE152
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2018
Initial Date Received by Manufacturer 12/08/2017
Supplement Date Received by Manufacturer04/05/2018
Initial Report FDA Received Date01/04/2018
Supplement Report FDA Received Date04/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-