| Model Number |
CQ7578 |
| Medical Device Problem Codes |
Break (1069); Peeled/Delaminated (1454); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No hospital/medical records or medical images have been made available to the manufacturer.The lot number for the device was not provided; therefore, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that after the first inflation/deflation attempt the pta balloon allegedly broke.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Conclusion: the device was returned for evaluation.A visual inspection found peeled pebax throughout the length of the balloon.Additionally, the balloon was inflated and a pinhole rupture was identified on the balloon.Therefore, the investigation is confirmed for a pinhole rupture as well as for peeled pebax on the balloon.The definitive root cause for the identified peeled pebax or pinhole rupture could not be determined based upon the available information.Labeling review: the current ifu (instructions for use) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter, introducer sheath and guidewire (as necessary) as a single unit.Use of the conquest pta dilatation catheter: while maintaining negative pressure and the position of the guidewire, grasp the balloon catheter just outside the sheath and withdraw the deflated dilatation catheter over the wire through the introducer sheath.Use of a gentle counterclockwise motion may be used to help facilitate catheter removal through the introducer sheath (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the brachial artery the pta balloon allegedly ruptured during the first inflation attempt (atm unknown).The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
|
| |
|