A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Unit passed the dat test at 0.08775 inches.All operating currents are within specification.No unexpected low battery or off no power alarms noted.Unit passed functional test including displacement test, rewind, basic occlusion, occlusion, prime/compromised force sensor system, and excessive no delivery alarm test.Unit functioned properly.No delivery anomaly noted during testing.Unit was received with minor scratched lcd window, cracked reservoir tube lip, cracked belt clip slot and broken battery tube thread.
|