MAUDE Adverse Event Report: AV-TEMECULA-CT MULTI-LINK RX VISION CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MULTI-LINK RX VISION CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1007849-15
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number -(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: a visual and dimensional inspection was performed on the returned stent implant.The reported stent dislodgement was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined that the reported stent dislodgement appears to be related to circumstances of the procedure due to inadvertent mishandling during protective sheath/stylet removal.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during preparation of the 3.5 x 15 mm rx vision stent delivery system (sds), when removing the protective sheath, the stent implant dislodged from the balloon.There was no report of resistance felt during removal of the sheath.The device could not be used on the patient.A new sds was used to complete the case successfully.There was no clinically significant delay in the procedure reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTI-LINK RX VISION CORONARY STENT SYSTEM
Common Device Name
CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7164988
Report Number2024168-2018-00067
Device Sequence Number19312264
Product Code MAF
UDI-Device Identifier08717648050701
UDI-Public08717648050701
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P020047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Catalogue Number1007849-15
Device Lot Number6110441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2017
Initial Report FDA Received Date01/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-