• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION HIP; OBSOLETE: ZIRCONIA HEAD 28 / NEU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION HIP; OBSOLETE: ZIRCONIA HEAD 28 / NEU Back to Search Results
Catalog Number 410-28-000
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 12/19/2017
Event Type  Injury  
Event Description
Revision surgery - due to the patient having an early generation 28 mm zirconia ceramic head which broke into pieces.The surgeon removed and replaced the head, liner, and added the sleeve.
 
Manufacturer Narrative
The reason for this revision surgery was due to the ceramic head breaking.The previous surgery and the revision detailed in this investigation occurred over 18 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the ceramic head breaking.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION HIP
Type of Device
OBSOLETE: ZIRCONIA HEAD 28 / NEU
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key7165050
MDR Text Key96393966
Report Number1644408-2017-01237
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K965002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/28/2004
Device Catalogue Number410-28-000
Device Lot Number468051
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/21/1999
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
431-28-008, LOT 428671
Patient Outcome(s) Required Intervention;
Patient Age85 YR
-
-