Model Number MMT-551NAS |
Device Problems
Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/26/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 194 mg/dl at the time of the incident.The customer reported that he was rewinding the pump because he was putting on a new set.When getting the alarm after there was insulin squirting out the infusion set.Customer stated that the drive support cap was sticking out.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to confirm compromised force sensor system alarm due to motor error alarm.Device received with corroded battery tube noted.No moisture damage on electronics and motor assembly noted.
|
|