| Model Number |
8888145044P |
| Medical Device Problem Code |
Hole In Material (1293)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the nurse was flushing the catheter with normal saline and noticed a tear in the catheter between the arterial connector and the silicone extension.The white plastic wrap around the silicone tube arterial connector was loose and could be rotated 360 degrees.The catheter was replaced and there was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: a device history record (dhr) review revealed no discrepancies that may have contributed to a complaint of this failure mode.All quality assurance testing performed during manufacturing was acceptable.All dhrs are reviewed for accuracy prior to product release.One catheter was received for analysis and investigation.Visual evaluation of the catheter revealed signs of use and a tear in the catheter between the arterial connector and silicone extension.As per the instructions for use (ifu), the customer must perform a visual inspection before using the device.It states to exercise caution when using sharp instruments near the catheter because the catheter tubing can tear when subjected to excessive force or rough edges.The reported condition was confirmed.Based on the sample received, the most probable root cause can be considered as damaged during use.The evidence provided is enough to discard the manufacturing process as a potential cause.No trends or triggers have been found.Therefore, no action is deemed necessary at this time.It must be noted that in-process controls are in place to prevent nonconforming product from leaving the manufacturing operations.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|