MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAB
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 12/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they had an unexpected restart.The customer¿s blood glucose level was 349 mg/dl.The customer stated that they had damage to their pump and was also getting an unexpected restart alarm.Customer declined troubleshoot for high blood glucose.The troubleshooting was performed and was advised to remove the current battery from the pump and insert a new battery (per ifu guidelines) and pump alarmed unexpected restart.The customer advised that the pump will need to be replaced.The customer advised to monitor the pump and that patient may continue to wear the pump until replacement arrives.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Pump passed the displacement test and unexpected restart error alarm test.No unexpected restart error alarm anomalies noted.Pump received with minor scratched lcd window, cracked case at the display window corners, cracked lcd window, cracked belt clip slot, broken battery tube threads, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7166592
Report Number3004209178-2018-41045
Device Sequence Number1656233
Product Code OYC
UDI-Device Identifier00643169077645
UDI-Public(01)00643169077645
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAB
Device Catalogue NumberMMT-523NAB
Device Lot NumberA7523NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2018
Initial Date Received by Manufacturer 12/27/2017
Supplement Date Received by Manufacturer05/24/2018
Initial Report FDA Received Date01/04/2018
Supplement Report FDA Received Date06/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age48 YR
Patient Weight114
-
-