MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/04/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The intraocular lens was removed and replaced during the same procedure.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that when the intraocular lens (iol) unfolded in the right eye, there was a full thickness break in it from the optic edge to the trailing haptic that the surgeon suspects happened when the iol was advanced in the cartridge.He had to cut the iol out and insert another one during the same procedure.The doctor used wound assisted technique.The procedure was completed successfully using a back-up lens.There was incision enlargement.No sutures required.No vitrectomy.The patient is stable.No additional information was provided to abbott medical optics.
 
Additional Manufacturer Narrative
Device available for evaluation? yes, returned to manufacturer on 1/19/2018.Device returned to manufacturer? yes.Device evaluation: the return product consisted of four pieces of the lens attached to gauze.Viscoelastic was observed on the pieces of lens returned.The insertion device was not returned; therefore, an evaluation was not possible.The reported issue was verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu provide the customer with proper usage instructions and guidelines.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7166899
Report Number2648035-2018-00018
Device Sequence Number1823743
Product Code HQL
UDI-Device Identifier05050474558175
UDI-Public(01)05050474558175(17)200911
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/11/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Initial Date Received by Manufacturer 12/06/2017
Supplement Date Received by Manufacturer01/23/2018
Initial Report FDA Received Date01/05/2018
Supplement Report FDA Received Date02/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-