• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 44MM M SPEC+1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 44MM M SPEC+1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 136561000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994)
Date of Event 12/05/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical der received that patient was revised to address adverse local tissue reaction.(b)(6) 2007.(b)(6) 2017.Right hip.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6 patient code: no code available (3191) used to capture the surgical intervention and blood heavy metal increased.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 44MM M SPEC+1.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key7167315
Report Number1818910-2018-50300
Device Sequence Number1475659
Product Code JDI
UDI-Device Identifier10603295034032
UDI-Public10603295034032
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number136561000
Device Lot Number2256445
Initial Date Received by Manufacturer 12/07/2017
Supplement Date Received by Manufacturer02/01/2018
09/19/2019
12/24/2019
Initial Report FDA Received Date01/05/2018
Supplement Report FDA Received Date02/02/2018
10/15/2019
01/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient Weight120
-
-