| Catalog Number |
299966321 |
| Medical Device Problem Codes |
Dull, Blunt (2407); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the mto pinnacle cup impactor was discovered at wash/sterilization to have the threaded tip insert extended from the face of the handle by approximately 1mm.A second handle was found with damaged threads.Both of these occurred after on use.Insert retaining pin appears to have sheared.No surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|