| Model Number |
H7493808015370 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device was not received for analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the left anterior descending artery.A 3.75mm x 15mm quantum¿ maverick¿ balloon catheter was advanced for dilatation and during inflation at 14-16 atmospheres, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The balloon was loosely folded.The tip is damaged.Microscopic examination revealed that the balloon has a pinhole 1.5mm from the proximal markerband.There is no indication the device was inflated over rated burst pressure.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the left anterior descending artery.A 3.75mm x 15mm quantum¿ maverick¿ balloon catheter was advanced for dilatation and during inflation at 14-16 atmospheres, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|