| Catalog Number |
FS-WL-O-S |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical devices: ocy, endoscope and accessories.Investigation evaluation: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all fusion wire guide locking devices are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
A medwatch report mw5073533 was received on 12/11/2017: the patient was undergoing endoscopic retrograde cholangiopancreatography (ercp) and it was noted that a sponge was in the biopsy channel from a previous procedure.It is believed that the sponge material machines [matches] that of the biopsy cap.The scope was removed and new scope was utilized.The patient required no further treatment.Additional information was received on 01/03/2017: the reporters were not certain if the locking device was from cook.It was quite possibly [the boston scientific] locking device because they use both.The reporter said she has sent all the information to risk management.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to common device name is being corrected to endoscopic guidewire, gastroenterology-urology and procode is being corrected to ocy.Investigation evaluation: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The user stated that a water soluble lubricant was not used.This could have possibly contributed to the wire lock becoming displaced.The instructions for use states, "apply water-soluble lubricant to top of cap as needed to ease passage of accessories." prior to distribution, all fusion wire guide locking devices are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
The following was initially provided on 01/05/2018: "a medwatch report mw5073533 was received on 12/11/2017: the patient was undergoing endoscopic retrograde cholangiopancreatography (ercp) and it was noted that a sponge was in the biopsy channel from a previous procedure.It is believed that the sponge material machines [matches] that of the biopsy cap.The scope was removed and new scope was utilized.The patient required no further treatment.Additional information was received on 01/03/2017: the reporters were not certain if the locking device was from cook.It was quite possibly [the boston scientific] locking device because they use both.The reporter said she has sent all the information to risk management.Additional information was received on 01/09/2018: the sponge was discovered in the biopsy channel.The endoscope was immediately removed from service and another ercp endoscope was used.
|
| |
|