Catalog Number 2030-6530-1
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Injury (2348); Ambulation or Postural Difficulties (2544)
Date of Event 12/11/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There has been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Not available.
 
Event or Problem Description
It was reported that patient's left hip was revised after patient complaint of pain.Intra-operatively, the surgeon reported that the 30mm screw was broken.Patient was revised to an mdm construct.A medical review by clinical consultant states "cup malposition in medialized position with additional adverse short leg length effects has contributed to an overload condition in the arthroplasty contributing to cup loosening requiring revision.".
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a bone screw was reported.The event was confirmed.Method & results: -device evaluation and results: not performed as device was not returned.-medical records received and evaluation: a review of the provided x-rays by a clinical consultant noted the following; [.] procedure-related factors: - cup malposition in medialized position.- major leg length difference, left hip short.Patient-related factors - no info.Device-related factors: - none.Diagnosis: - cup malposition in medialized position with additional adverse short leg length effects has contributed to an overload condition in the arthroplasty contributing to cup loosening requiring revision.[.] -device history review: review indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information, including patient details, operative reports and return of device are needed to fully investigate the event.If further information becomes available this investigation will be re-opened.
 
Event or Problem Description
It was reported that patient's left hip was revised after patient complaint of pain.Intra-operatively, the surgeon reported that the 30mm screw was broken.Patient was revised to an mdm construct.A medical review by clinical consultant states "cup malposition in medialized position with additional adverse short leg length effects has contributed to an overload condition in the arthroplasty contributing to cup loosening requiring revision.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5 CANCELLOUS BONE SCREW 30MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7167968
Report Number0002249697-2018-00058
Device Sequence Number13816779
Product Code HWC
UDI-Device Identifier07613327037005
UDI-Public07613327037005
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K894124
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/15/2021
Device Catalogue Number2030-6530-1
Device Lot Number3R5DM6
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2017
Supplement Date Received by Manufacturer02/01/2018
Initial Report FDA Received Date01/05/2018
Supplement Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
-
-