Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
 
Event or Problem Description
On (b)(6) 2017, the reporter contacted animas, alleging a power (damage) issue.It was noted that the battery compartment was cracked.There was no allegation of an adverse event associated with this complaint.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 15-jan-2018 with the following findings: a review of the pump black box data showed unexplained pump reboots.During visual inspection of the pump, it was observed that the battery compartment was cracked in two places.The returned battery cap had stripped threads and was unable to fully attach to the pump; the pump reboots were duplicated with the returned battery cap.A new test battery cap was able to carefully attach to the pump and was used to complete a 24 hour duration test and it.There were no pump reboots duplicated during this time with the test battery cap.The pump cover was removed and there was no evidence of moisture or loose components inside pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7168207
Report Number2531779-2018-00358
Device Sequence Number1475678
Product Code OYC
UDI-Device Identifier10840406100020
UDI-Public10840406100020
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 12/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2018
Was the Report Sent to FDA? Yes
Device Age22 MO
Initial Date Received by Manufacturer 12/11/2017
Supplement Date Received by Manufacturer12/11/2017
Initial Report FDA Received Date01/05/2018
Supplement Report FDA Received Date02/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
-
-