• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE Back to Search Results
Catalog Number 1009665
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Concomitant medical devices: emerge 3mm x 20mm; chocolate balloon 3.5mm x 15mm.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It should be noted the hi torque guide wire instructions for use (ifu) states: this hi-torque guide wire is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (ptca) and percutaneous transluminal angioplasty (pta).This guide wire may also be used with compatible stent devices during therapeutic procedures.The investigation was unable to determine a conclusive cause for the reported difficulties of difficult to position resulting in a core separation.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to perform a left popliteal bypass graft.A 300 cm balance middleweight universal ii guide wire was used and met resistance with a non-abbott balloon catheter which was inside the anatomy during advancement.Additionally, the core of the guide wire separated inside the balloon catheter.The separation occurred outside the anatomy; therefore, both devices were simply withdrawn from the anatomy.Then, the devices were replaced with a 3.0 x 20 mm and a 3.5 x 15 mm non-abbott balloon catheter to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7168350
Report Number2024168-2018-00100
Device Sequence Number933031
Product Code DQX
UDI-Device Identifier08717648118722
UDI-Public08717648118722
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K072460
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number1009665
Device Lot Number7060671
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2017
Initial Report FDA Received Date01/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight100
-
-