| Medical Device Problem Code |
Failure to Prime (1492)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a prime (loss of prime) issue.There was no indication that the issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: device evaluation: the device has been returned and evaluated by product analysis on 11-jan-2018 with the following findings: there were multiple loss of prime warnings observed in the pump's black box.The pump was exercised for 24 hours with no loss of prime warnings observed.The force sensor passed a calibration check with no issues.
|
| |
|