| Model Number |
6290 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that the communicator was not working and had no lights.A boston scientific company representative verified that there was a recent electrical outage and also verified that the power cord has a steady light.The patient observed that there was a burning smell and the communicator was warm to the touch.The company representative had the patient unplugged the power cord from the power outlet.The company representative then ordered a communicator replacement and a return label was sent.Furthermore, the company representative verified that the patient had a new address.The communicator was deactivated.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, the allegation against the communicator was not confirmed as the communicator had pest infestation.The product was returned in a condition that made analysis impossible therefore analysis was unable to confirm the observations.
|
| |
|