| Model Number |
LUN3012-08 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Unintended Movement (3026); Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
08/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Post-operative pain and implant migration are listed in the device instructions for use (ifu) and in the procedural technique guide as known possible adverse effects associated with use of the system.A definitive root cause for this device failure could not be determined due to device not being returned for investigation.However, based on the imaging, the intraoperative factors that may have contributed to the implant migration include: luna footprint was not a full 360 deployment; middle component was not fully seated or interlocked with the outer components; enough micro-motion of the segment occurred to cause the middle component to migrate; and the lock wire eventually failed under the increased strain from movement of the middle component, and that failure allowed the middle to migrate even further posteriorly and into the canal.A review of the lot history record confirmed no manufacturing nonconformities issued to the reported lot that would have caused or contributed to this event.Based on review of the information provided, there is no indication of a product quality issue with respect to manufacture, design or labeling.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the company became aware of a revision surgery performed that day due to patient experiencing severe pain.It was further learned that the original case was completed on (b)(6) 2017.On (b)(6) 2017, 2-month post-op, the patient presented in the er with severe pain down his right leg.Ct scan showed that a portion of the implant had migrated posteriorly into the spinal canal.The surgeon scheduled a revision surgery on the following day to remove the implant.During the revision surgery, the implant could not be safely removed as it was too medial and under the dura.The surgeon performed foraminotomies to create more space for the nerve roots.Patient is doing okay one day post-op follow-up.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as a supplemental report to provide additional information.On august 16, 2019, benvenue completed an investigation of the removed implant that was retrieved on april 18, 2018.The investigation confirmed the conclusions of the initial investigation, which relied on images alone; these were: incomplete deployment of implant and middle component not fully interlocked with outer components.In addition, enough micro-motion of the segment occurred to cause the middle component to migrate and the lock wire eventually failed under the increased strain from movement of the middle component.This allowed the middle to migrate even further posteriorly and into the canal.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as a supplemental report to provide additional information.On may 14, 2019, a former benvenue regional sales manager reported that he had been contacted by an attorney representing the patient in this case.The following day, may 15, 2019, through a phone conversation with the former benvenue regional sales manager, benvenue became aware that the patient had undergone a second revision surgery that resulted in explantation of the device.Benvenue's attorney is contacting the medical university of (b)(6) to request return of the implant for investigation.A follow-up mdr report will be filed if benvenue is successful in obtaining and investigating the implant.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as a supplemental report to provide additional information.On may 15, 2019, through a phone conversation with the former sales manager, benvenue became aware that the patient had undergone a second revision surgery that resulted in explantation of the device.The company is requesting return of the implant for investigation.A follow-up mdr report will be filed if benvenue is successful in obtaining and investigating the implant.
|
| |
|