WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12MM/130 DEG TI CANN TFNA 400MM/RIGHT-STERILE; ROD,FIXATION,INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.037.260S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: date of event is unknown.Date of implant is unknown.Approximately six months prior to the revision on (b)(6) 2017.(b)(4).The device was received and the product evaluation is in progress.No conclusion can be drawn.Device history records review pending completion.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a revision of trochanteric fixation nail advanced (tfna) on (b)(6) 2017 due to non-union, broken tfna nail and broken tfna distal locking screw.The patient is reported to have been non-compliant and walked without canes or crutches as opposed to the post-operative instructions for the first three months after the initial procedure that was conducted about six months prior to the revision surgery.The tfna nail and distal locking screw were observed to be broken during a follow-up x-ray after initial surgery.Nail was broken just proximal to the lag screw insertion hole.The removed hardware included one broken tfna nail, one intact lag screw and one broken 5.0 mm distal locking screw.The patient was revised to unknown quantity of 95-degree condylar blade plates and 4.5 mm cortical screw construct on the proximal femur.Allografts and autografts were used to treat the nonunion.The lag screw site was filled with norian.During the revision surgery, the surgeon used a helical blade coupling screw into the distal portion of the nail so as to remove the distal part of the nail from the patient.During the removal of this portion of the nail, the mallet slipped off the driving cap and struck the proximal end of the coupling screw, thus damaging the secondary threads of the coupling screw that goes inside a t-handle inserter.The surgeon then again tried inserting the coupling screw into the t-handle inserter, thus damaging the t-handle inserter as well.The damaging of the t-handle inserter is reported to have had no effect on the surgery.Procedure was completed without any complications and without any surgical delay.This complaint captures the post-operative malfunction of the devices.(b)(4) captures the intraoperative incident of revision surgery.Concomitant devices reported: lag screw (part # 04.038.100s, lot # 9909007, quantity # 1).This report is for one (1) 12mm/130 deg ti cann tfna 400mm/right ¿ sterile.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record review was performed for the subject device: manufacturing location: (b)(4).Manufacturing date: 26-feb-2016, expiration date: 31-jan-2026.Part #: 04.037.260s, lot#: h043062 (sterile) - 12mm/130 deg ti cann tfna 400mm/right - sterile.Quantity 6.Component parts reviewed: 04.037.912.4 - wave spring, shim ended bp-55 lot - 7921061 04.037.942.2 - lock prong 130 degree, tfna bp-55 lot - 9667976 04.037.912.3 - tfna lock drive bp-58 lot ¿ 9981088 21127 - raw material lot bp-80 lot - 7815682.Raw material was received from (b)(4).Certificate of analysis received for (b)(4) from (b)(4) meet specification.Raw material receiving/putaway checklist meet requirements.Inspection sheet for in-process/inspect dimensional/final met inspection acceptance criteria.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Scn no: ¿sterility documentation was reviewed and determined to be conforming.¿ no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.A product development investigation was performed for the subject device: the returned 12mm/130 degree ti cannulated tfna 400mm/left (04.037.260s) is a component of the titanium tfna system for intramedullary fixation of proximal femur fractures.The returned device was examined and the complaint condition was able to be confirmed as the device was found to be broken at the proximal locking hole.Based on the complaint description the failure occurred as a result of patient non-compliance related to load-bearing.The complaint condition was unable to be replicated due to post-manufacturing damage.Relevant drawings for the returned device were reviewed (both current revision and from the time of manufacture): the design, materials and finishing processes were found to be appropriate for the intended use of this device.No dimensional analysis is possible due to post-manufacturing damage.A device history review, including material and hardness reviews, was performed for the returned implant¿s lot number and no mrrs, ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|