A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Unit passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime test, excessive no delivery test and dat test at 0.08740 inches.The stop (idle) current and run current measurement tests are within specification.Unit also passed self test, off no power alarm test and error test.No battery out limit alarm, failed battery test alarm, alarm or alarm noted during testing.No unexpected no delivery alarm noted.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No bolus anomaly or basal anomaly noted during testing.Unit had minor scratched display window, debris under the display window, cracked case at display window corner, cracked battery tube threads, cracked reservoir tube, cracked reservoir tube window, a partially broken off reservoir tube lip and a stained end cap sticker.
|