• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR (BPV) SIDEKICK SUPPORT CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR (BPV) SIDEKICK SUPPORT CATHETER Back to Search Results
Catalog Number SD70ATG
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was not returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during treatment of a lesion site in the popliteal, the recanalization catheter allegedly became stuck in the support catheter's sheath.It was further reported that the recanalization catheter and the support catheter were allegedly removed.Reportedly, another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the investigation was inconclusive for retraction problem as the device was not returned.The root cause could not be determined based upon available information.It was unknown whether patient and/or procedural factors contributed to the event.Labeling review: the current ifu (instructions for use) state: product length selection and procedural notes: when using the tapered sidekick catheter with the crosser catheter 14s or 14p, the crosser catheter can be advanced approximately 15cm from the tip of the sidekick catheter.Resistance is encountered beyond 15cm due to the taper on the crosser catheter aligning with the taper on the sidekick catheter.A taper lock-up marker (single marker on the crosser catheter shaft) is located 127cm from the distal tip for the 146cm crosser catheter and 80cm from the distal tip for the 106cm crosser catheter.The taper lock-up marker is an indicator that the tapers on the catheters are nearing alignment; advance the crosser catheter slowly.Do not continue to advance the crosser catheter if resistance is encountered.Procedural steps: remove product using standard techniques while maintaining guidewire access when use of the product is complete.
 
Event or Problem Description
It was reported that during treatment of a lesion site in the popliteal, the recanalization catheter allegedly became stuck in the support catheter's sheath.It was further reported that the recanalization catheter and the support catheter were allegedly removed.Reportedly, another device was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIDEKICK SUPPORT CATHETER
Common Device Name
SUPPORT CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR (BPV)
KS 
MDR Report Key7170139
Report Number2020394-2017-01888
Device Sequence Number16617055
Product Code DQY
UDI-Device Identifier10801741125611
UDI-Public(01)10801741125611
Combination Product (Y/N)N
PMA/510(K) Number
K161986
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2020
Device Catalogue NumberSD70ATG
Device Lot NumberGFBR3676
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2047
Supplement Date Received by Manufacturer01/29/2018
Initial Report FDA Received Date01/05/2018
Supplement Report FDA Received Date01/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-