• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Codes Difficult to Insert (1316); Unintended Movement (3026)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
Date of Event 11/30/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lens was inserted and removed.(b)(4).An attempt has been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a pcb00 26.0 diopter intraocular lens (iol) deployed itself upside down and rotated.There was also a capsular tear that occurred and a vitrectomy was performed.The lens was then removed and the patient was left aphakic.Thereafter, the patient was referred to another hospital.Reportedly, the technician stated she may have inserted the lens on the wrong side.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product was not returned for investigation as it was discarded by the customer.The reported complaint was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no other complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu provide the customer with proper usage instructions and guidelines.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7170413
Report Number2648035-2018-00046
Device Sequence Number1773288
Product Code HQL
UDI-Device Identifier05050474558380
UDI-Public(01)05050474558380(17)200502
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date05/02/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000260
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/07/2017
Supplement Date Received by Manufacturer01/15/2018
Initial Report FDA Received Date01/05/2018
Supplement Report FDA Received Date02/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-