Model Number 030453
Medical Device Problem Codes Misfire (2532); Failure to Form Staple (2579)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was previously reported to the fda on a summary report and is now being submitted on a 3500 a per fda request.Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the cartridge revealed that the loading unit was partially fired.The anvil of the loading unit was bowed and the knife bar assembly was buckled.Microscopic examination of the loading unit observed damaged to the cutting edge of the knife blade.Additionally, the staple cartridge was disengaged from the channel.Due to the noted damage no functional testing was performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the damaged knife blade and disengaged staple cartridge may occur when an obstacle has been incorporated in the jaws during application.The information booklet which accompanies each product shipment cautions the user; ensure that no obstructions (such as clips) are incorporated in the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Replication of the bowed anvil, deformed anvil clamping mechanism, and buckled knife-bar assembly may occur under the following conditions firing over tissue that is beyond the recommended thickness range.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly and tissue may not be fully transected.The information booklet which accompanies each product shipment offers the following as a warning and precaution." preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the laparoscopic appendectomy procedure the device partially fired, there was poor staple formation and the staple line was incomplete at the distal end.To complete the procedure, a new device was used.There was no harm caused to the patient.Current patient status is alive with no injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7170900
Report Number1219930-2018-00185
Device Sequence Number19093850
Product Code GDW
UDI-Device Identifier20884523002840
UDI-Public20884523002840
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 01/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model Number030453
Device Catalogue Number030453
Device Lot NumberN6K0442KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2017
Initial Report FDA Received Date01/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No