This event was previously reported to the fda on a summary report and is now being submitted on a 3500 a per fda request.Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the cartridge revealed that the loading unit was partially fired.The anvil of the loading unit was bowed and the knife bar assembly was buckled.Microscopic examination of the loading unit observed damaged to the cutting edge of the knife blade.Additionally, the staple cartridge was disengaged from the channel.Due to the noted damage no functional testing was performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the damaged knife blade and disengaged staple cartridge may occur when an obstacle has been incorporated in the jaws during application.The information booklet which accompanies each product shipment cautions the user; ensure that no obstructions (such as clips) are incorporated in the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Replication of the bowed anvil, deformed anvil clamping mechanism, and buckled knife-bar assembly may occur under the following conditions firing over tissue that is beyond the recommended thickness range.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly and tissue may not be fully transected.The information booklet which accompanies each product shipment offers the following as a warning and precaution." preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
|
According to the reporter, during the laparoscopic appendectomy procedure the device partially fired, there was poor staple formation and the staple line was incomplete at the distal end.To complete the procedure, a new device was used.There was no harm caused to the patient.Current patient status is alive with no injury.
|