| Model Number |
4078-9 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned cable was evaluated.During evaluation the cable passed all visual and functional testing.The cable was determined to be functioning as designed.
|
| |
|
Event or Problem Description
|
|
The customer reported: "was told by rn (b)(6) that the monitor in room 448 sporadically loses the spo2 reading.She changed cables twice during her shift to get it to work.I have the 2 cables from the specific room but not any sensors.As tasked by leadership, i am to pull the cable and send it for failure analysis.Unfortunately the rn did not save the sensor.Logs were pulled and i didn't see any dropouts, just large desats." no patient impact or consequences were reported.
|
| |
|