| Catalog Number |
LR-EVN-13.0-RL |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Perforation of Vessels (2135)
|
| Date of Event |
12/12/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The pma/510(k) # - k141148.Device - the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Event is currently under investigation.A follow up report will be submitted upon receipt of additional information or completion of investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during the course of the lead extraction procedure, a small bleed occurred in a minor branch of the innominate vein.Intervention was required in the form of a chest drain and both leads were extracted successfully.According to the initial reporter, after putting the drain in and continuing with the lead extraction, "patient is fine." the leads had been implanted for over 17 years.No additional patient consequences were reported.
|
| |
|
Event or Problem Description
|
|
See additional mfr narrative for the investigation of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # - k141148.Investigation summary: the affected product was not returned for this complaint and therefore a physical investigation could not be performed.The instructions for use was reviewed and listed as a potential adverse event is 'laceration or tearing of vascular structures or the myocardium'.This complaint event is a known failure mode and will be monitored and trended through the complaint handling and post market processes.A device history record review was performed including the manufacturing and quality control records.All steps were documented per procedure by trained personnel and there were no signs to indicate that a nonconforming device was released to the field.This complaint could only be confirmed through customer testimony.
|
| |
|