| Catalog Number |
NC3-024115 |
| Medical Device Problem Codes |
Kinked (1339); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The event is currently under investigation.A follow up report will be submitted upon receipt of additional information or completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported the ncompass nitinol stone extractor wire kinked approximately 3cm from the basket handle, preventing operation of the basket.No unintended portion of the device remained inside the patient¿s body.No additional procedures were required due to this occurrence.No adverse effects or consequences were reported to the patient due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unavailable or unchanged.Investigation ¿ evaluation.(b)(4).A review of manufacturing instructions, specifications, quality control data, documentation, visual inspection, and functional testing was conducted during the investigation.A document-based investigation evaluation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.The complaint device was returned for evaluation.The device was found to be non-functional due to a severe sheath kink at the support tube.The ifu contains a caution to not use excessive force to manipulate the device, or damage to the device may occur.It is unknown if excessive force was used on the device.Based on the provided information a definitive root cause cannot be established or reported at this time.We will notify the appropriate personnel and continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
|
| |
|
Search Alerts/Recalls
|