| Model Number |
I7 |
| Medical Device Problem Codes |
No Display/Image (1183); Display or Visual Feedback Problem (1184); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/11/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was unavailable from the site.A medtronic representative inspected the navigation system onsite and confirmed the issue.The representative was able to identify the touch panel input (input #2) on the transceiver was related to the monitor presenting the problem.The transceiver touch panel input #2 had a fiber optic cable labeled channel 1 connected and the monitor the fiber optic cable connected was labeled #2.The issue was resolved by changing the fiber optic cable at the monitor to channel #1.No parts required replacement.No parts were returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that while in a procedure the touch screen monitor on the navigation system showed "searching for dvi connection".The monitor is not functioning.The site used the other monitors and proceeded with the case.There was no delay to the procedure as a result of this issue and no impact on the patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: patient demographics provided.A medtronic representative reported that the issue occurred in an oblique lumbar interbody fusion (olif).
|
| |
|
Additional Manufacturer Narrative
|
|
Device code was corrected.If information is provided in the future, a supplemental report will be issued.
|
| |
|